Project Description
Product Features:
The instrument is controlled by a large-capacity single-chip microcomputer, and the work is stable and reliable
A wide-range watchdog circuit is installed in the instrument to prevent crashes
A variety of operation options, the instrument program has two national standard methods GB507-1986 and GB507-2002 and custom operations, which can adapt to multiple choices of different users
The oil cup of the instrument is made of special glass by one-time casting to prevent the occurrence of interference phenomena such as oil leakage.
High measurement accuracy, strong anti-interference ability, safe and reliable
DH801 series insulating oil withstand voltage tester is our company’s all scientific research and technical personnel,it also call the Dielectric strength tester , according to the relevant regulations of the national standard GB507-1986 and the industry standard DL-474·4-92DL/T596-1996.
Exerting its own advantages, after many on-site tests and Long-term unremitting efforts, meticulously researched and developed high-accuracy, fully digital industrial instruments. To meet the needs of different users.
The instrument oil cup is made of special glass by one-time casting to prevent the occurrence of interference phenomena such as oil leakage;
The unique high-voltage end sampling design of the instrument allows the test value to directly enter the A/D converter, avoiding the error caused in the analog circuit, and making the measurement result more accurate;
The instrument has functions such as over-current, over-voltage, short-circuit protection, etc., and has strong anti-interference ability and good electromagnetic compatibility;
Portable structure, easy to move, very convenient for indoor and outdoor use.
1. Booster capacity | 1.5kVA |
---|---|
2. Boost speed | 2.0 kV/s, 3.0 kV/s, optional second gear |
Error | 0.2kV/s |
3. Output voltage | 0~80 kV |
4. Power distortion rate | <1% |
5. Display mode | Large-screen LCD Chinese character display |
6. Electrode gap | Standard 2.5 mm |
7. Overall dimensions | Length 430 mm; width 350 mm; height 370 mm |
8. Instrument weight | 36 kg |
Conditions of Use | |
1. Ambient temperature | 0~40℃ |
2. Relative humidity | ≤ 85% |
3. Working power supply | AC 220V ± 20% |
4. Power frequency | 50 Hz ± 5 Hz |
5. Power consumption | <200 W |