Project Description

Product Features:

  • The instrument is controlled by a large-capacity single-chip microcomputer, and the work is stable and reliable

  • A wide-range watchdog circuit is installed in the instrument to prevent crashes

  • A variety of operation options, the instrument program has two national standard methods GB507-1986 and GB507-2002 and custom operations, which can adapt to multiple choices of different users

  • The oil cup of the instrument is made of special glass by one-time casting to prevent the occurrence of interference phenomena such as oil leakage.

  • High measurement accuracy, strong anti-interference ability, safe and reliable

DH801 series insulating oil withstand voltage tester is our company’s all scientific research and technical personnel,it also call the Dielectric strength tester , according to the relevant regulations of the national standard GB507-1986 and the industry standard DL-474·4-92DL/T596-1996.

Exerting its own advantages, after many on-site tests and Long-term unremitting efforts, meticulously researched and developed high-accuracy, fully digital industrial instruments. To meet the needs of different users.

The instrument oil cup is made of special glass by one-time casting to prevent the occurrence of interference phenomena such as oil leakage;
The unique high-voltage end sampling design of the instrument allows the test value to directly enter the A/D converter, avoiding the error caused in the analog circuit, and making the measurement result more accurate;
The instrument has functions such as over-current, over-voltage, short-circuit protection, etc., and has strong anti-interference ability and good electromagnetic compatibility;
Portable structure, easy to move, very convenient for indoor and outdoor use.

1. Booster capacity1.5kVA
2. Boost speed2.0 kV/s, 3.0 kV/s, optional second gear
Error0.2kV/s
3. Output voltage0~80 kV
4. Power distortion rate<1%
5. Display modeLarge-screen LCD Chinese character display
6. Electrode gapStandard 2.5 mm
7. Overall dimensionsLength 430 mm; width 350 mm; height 370 mm
8. Instrument weight36 kg
Conditions of Use
1. Ambient temperature0~40℃
2. Relative humidity≤ 85%
3. Working power supplyAC 220V ± 20%
4. Power frequency50 Hz ± 5 Hz
5. Power consumption<200 W
Fill out my online form.